Captus® 4000e Thyroid Uptake System, all models
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Captus® 4000e Thyroid Uptake System, all models |
Serial number 940000 to 940535 |
5430-30151 |
Issue
In 2023, three Captus® 4000e systems incurred a spring arm failure, causing the arm to fall. The collimator and arm, together, weigh 45 lbs (20.4 kg), which carries a risk of injury if the collimator comes into contact with the technologist or the patient.
Recall start date: November 26, 2023
Additional information
Details
Mirion Technologies (Capintec), Inc.
7 Vreeland Road, Florham Park, New Jersey, United States, 07932