CALiaGold PierceTube
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
CALiaGold PierceTube |
40045 |
1151300 |
Issue
Following a customer complaint, Sentinel detected a defect in three (3) lots of the collection tube CALiaGold PierceTube, ref. 1151300: lot 31390, 40037, 40045. The defect is an inner plastic crack that connects the two chambers of the collection tube, causing a reduction of the buffer volume needed for the measurement. It cannot be excluded that a significant volume decrease may lead to the impossibility to perform the test. The root cause of the plastic crack is still under investigation.
Recall Start Date: June 13, 2024