Health product recall

CADD-Solis Ambulatory Infusion Pump - Battery Pack

Last updated

Summary

Product
CADD-Solis Ambulatory Infusion Pump - Battery Pack
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

CADD-Solis Ambulatory Infusion Pump - Battery Pack

All lots.

21-2160-02
21-2160-51

CADD-Solis Power Cord/Batteries

All lots.

21-2160-51

Issue

Smiths Medical identified three (3) reports in which damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor may potentially lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.

Recall start date: July 22, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Smiths Medical ASD, Inc.

6000 Nathan Lane N., Minneapolis, Minnesota, United States, 55442

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75942

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe