Cadd Medication Cassette Reservior
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Cadd Medication Cassette Reservior | More than 10 numbers, contact manufacturer. | 21-7301-24 21-7300-24 21-7302-24 |
Issue
There is a potential that Cadd-legacy pumps may not detect that 50 ml and 100 ml Cadd medication cassette reservoirs with flow stop are attached to the pump when the cassettes are properly attached. This issue does not impact 250 ml flow-stop and non-flow stop Cadd medication cassette reservoirs. Manufacturing variations on certain cadd medication cassette reservoirs with flow stop may interfere with the pump detecting a properly attached cadd cassette. In such situations, the Cadd-legacy pump will issue a no disposable attached (NDA) double-beep audible warning if the pump cannot determine that the Cadd cassette is properly attached. The pump will initiate an NDA alarm if the nda double-beep warning is not resolved within 2 minutes. The user must clear the alarm and resolve the cause of the nda event before using the pump.
Recall start date: November 29, 2022
Additional information
Details
Smiths Medical Asd, Inc.
6000 Nathan Lane N., Minneapolis, Minnesota, United States, 55442
Get notified