Health product recall

Cadd Medication Cassette Reservior

Last updated

Summary

Product
Cadd Medication Cassette Reservior
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Cadd Medication Cassette Reservior

More than 10 numbers, contact manufacturer.

21-7301-24
21-7300-24
21-7302-24

Issue

There is a potential that Cadd-legacy pumps may not detect that 50 ml and 100 ml Cadd medication cassette reservoirs with flow stop are attached to the pump when the cassettes are properly attached. This issue does not impact 250 ml flow-stop and non-flow stop Cadd medication cassette reservoirs. Manufacturing variations on certain cadd medication cassette reservoirs with flow stop may interfere with the pump detecting a properly attached cadd cassette. In such situations, the Cadd-legacy pump will issue a no disposable attached (NDA) double-beep audible warning if the pump cannot determine that the Cadd cassette is properly attached. The pump will initiate an NDA alarm if the nda double-beep warning is not resolved within 2 minutes. The user must clear the alarm and resolve the cause of the nda event before using the pump.

Recall start date: November 29, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Smiths Medical Asd, Inc.

6000 Nathan Lane N., Minneapolis, Minnesota, United States, 55442

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72398