This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

CADD Blue-Striped Administration Set (2017-09-18)

Starting date:
September 18, 2017
Posting date:
October 20, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64900

Affected products

CADD Blue-Striped Administration Set

Reason

A detailed investigation points to a unique use case, involving a single Canadian customer using non-approved Smiths Medical Pouches, with the CADD Blue-Striped Administration Set. Use of non-approved accessories, as indicated in labeling, may adversely affect pump operation. Smiths Medical has decided to proceed with sending communication to notify all Canadian customers who have purchased the CADD Blue-Striped Administration Set.

Affected products

CADD Blue-Striped Administration Set

Lot or serial number

More than 10 numbers, contact manufacturer.  

Model or catalog number

21-7036-01

Companies
Manufacturer
Smiths Medical ASD, Inc.
1265 Grey Fox Road
St. Paul
55112
Minnesota
UNITED STATES