CADD Blue-Striped Administration Set (2017-09-18)
- Starting date:
- September 18, 2017
- Posting date:
- October 20, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-64900
Affected products
CADD Blue-Striped Administration Set
Reason
A detailed investigation points to a unique use case, involving a single Canadian customer using non-approved Smiths Medical Pouches, with the CADD Blue-Striped Administration Set. Use of non-approved accessories, as indicated in labeling, may adversely affect pump operation. Smiths Medical has decided to proceed with sending communication to notify all Canadian customers who have purchased the CADD Blue-Striped Administration Set.
Affected products
CADD Blue-Striped Administration Set
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
21-7036-01
Companies
- Manufacturer
-
Smiths Medical ASD, Inc.
1265 Grey Fox Road
St. Paul
55112
Minnesota
UNITED STATES