Health product recall

Cadd Administration Set

Last updated

Summary

Product
Cadd Administration Set
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Medication Cassett Reservior

More than 10 numbers, contact manufacturer.

21-7308-24
21-7301-24
21-7309-24
21-7300-24
21-7310-24
21-7302-24

Cadd Yellow Administration Set With Nrfit Connectors

More than 10 numbers, contact manufacturer.

21-7624-24
21-7649-24

Cadd Administration Set

More than 10 numbers, contact manufacturer.

21-7321-24
21-7324-24
21-7349-24
21-7395-24
21-7339-24
21-7343-24
21-7391-24
21-7333-24
21-7394-24
21-7346-24
21-7323-24
21-7336-24
21-7363-24
21-7383-24
21-7390-24
21-7322-24
21-7359-24

Issue

Manufacturing variations may cause the green cadd flow stop arm to compress and partially occlude the tubing before clinical use. If this occurs, there is a potential that the occlusion does not resolve when an affected reservoir or administration set is connected to the pump, and the pump may not detect the occlusion. This may result in underdelivery, despite the pump displaying that the infusion is running properly. Cadd flow stop medication cassette reservoirs and cadd flow stop administration sets provide free-flow protection. by design, a green, spring-loaded pivoting arm automatically squeezes the tubing closed when the reservoir or administration set is not installed on a pump. Connecting the reservoir or administration set to the pump causes the pump to push the flow stop arm, enabling fluid flow through the tubing. In certain circumstances, the tubing may remain occluded even though the cadd reservoir or administration set is loaded into the pump.

Recall start date: November 29, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Smiths Medical Asd, Inc.

6000 Nathan Lane N., Minneapolis, Minnesota, United States, 55442

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72397

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe