This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

CADD Administration Set

Starting date:
March 17, 2017
Posting date:
April 13, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62996

Reason

A drawing on the IFU incorrectly represents the actual filter size of the contents. Where the drawing should indicate that the package contains an administration set with a 0.2 micron filter, it is incorrectly illustrated as a 2.0 micron filter. The actual product in the package and the text based description on the IFU correctly conformed to the 0.2 micron filter specification for P/N 21-7339-24. It was the filter label on the drawing that was incorrect.

Affected products

CADD Administration Set

Lot or serial number

  • 46x1001
  • 46x817
  • 46x828
  • 46x829
  • 46x831
  • 46x896
  • 46x897
  • 46x899
  • 46x990

Model or catalog number

21-7339-24

Companies

Manufacturer

Smiths Medical ASD, Inc.

1265 Grey Fox Road

St. Paul

55112

Minnesota

UNITED STATES