This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
C-450 Fixed Ceiling Lift
- Starting date:
- August 22, 2012
- Posting date:
- November 26, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Product Safety, Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-15912
Recalled Products
- C-450 Fixed Ceiling Lift
Reason
An incident occurred at a facility where a patient dropped 6" into a commode. There were no injuries reported. The lift was returned to Waverley Glen and they determined that the C-450 lift was missing two C-clips at the end of the shaft.
Affected products
A. C-450 Fixed Ceiling Lift
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
9890 010 8390X
Companies
- Manufacturer
-
Prism Medical USA