This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

BUCKY DIAGNOST (2015-07-19)

Starting date:
July 19, 2015
Posting date:
August 18, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54682

Affected Products

A. BUCKY DIAGNOST

Reason

The trauma supporting block (connecting the trauma arm with the ceiling suspension telescope) is fixed with 3 screws, which are also used to adjust the correct alignment of it. Philips has become aware that in case this is not done in the correct way, the trauma arm is not fastened securely anymore and could fall down.

Affected products

A. BUCKY DIAGNOST

Lot or serial number

394898/SN08000161

Model or catalog number

9890 010 83821

Companies
Manufacturer
Philips Medical Systems Dmc Gmbh
Rontgenstrabe 24
Hamburg
22335
GERMANY