BUCKY DIAGNOST (2015-07-19)
- Starting date:
- July 19, 2015
- Posting date:
- August 18, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54682
Affected Products
A. BUCKY DIAGNOST
Reason
The trauma supporting block (connecting the trauma arm with the ceiling suspension telescope) is fixed with 3 screws, which are also used to adjust the correct alignment of it. Philips has become aware that in case this is not done in the correct way, the trauma arm is not fastened securely anymore and could fall down.
Affected products
A. BUCKY DIAGNOST
Lot or serial number
394898/SN08000161
Model or catalog number
9890 010 83821