Brivo NM 615 - Main Unit
Last updated
Summary
Product
Brivo NM 615 - Main Unit
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected product |
Lot or serial number |
Model or catalogue number |
---|---|---|
Brivo NM 615 - Main Unit |
24038 |
BRIVO NM 615 |
Issue
GE Healthcare has become aware that certain nuclear medicine systems could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms.
Recall start date: December 16, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
GE Hualun Medical Systems Co. Ltd. |
No.1 Yong Chang North Road, Beijing, Beijing, China, 100176 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76774
Get notified
Receive notifications for new and updated recalls and alerts by category.