This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Brilliance CT Big Bore System-Main Unit and Ingenuity CT System (2014-12-10)

Starting date:
December 10, 2014
Posting date:
December 23, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43005

Recalled Products

  1. Brilliance CT Big Bore System-Main Unit
  2. Ingenuity CT System

Reason

A potential problem with the vertical motor/brake assembly on the patient support/couch was discovered with a specific patient support/couch manufacturing lot, which may result in uncommanded vertical patient support/couch motion. To date, there have been no reported occurrences of harm.

Affected products

A) Brilliance CT Big Bore System-Main Unit

 

Lot or serial number

All lots.

Model or catalog number

4535, 670, 88051

Companies
Manufacturer

Philips Medical Systems (Cleveland), Inc.

595 Miner Road

Cleveland

44143

Ohio

UNITED STATES



B) Ingenuity CT System

Lot or serial number

All lots.

Model or catalog number

728323

Companies
Manufacturer

Philips Medical Systems (Cleveland), Inc.

595 Miner Road

Cleveland

44143

Ohio

UNITED STATES