Health product recall

Breeze C3/C4 Mobility Scooters (2013-11-11)

Starting date:
November 11, 2013
Posting date:
November 29, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36905

Recalled Products

A. Breeze C3 Mobility Scooter
B. Breeze C4 Mobility Scooter

Reason

Potential scenario where an internal potentionmeter wiper could become disconnected due to extensive wear on a certain number of assembled devices. If this were to occur the device could continue to drive forward at a slow speed of 2.4 mph and the only way to stop the device would be by turning off the main ignition key.

Affected products

A.     Breeze C3 Mobility Scooter
 

Lot or serial number

More than 100 numbers, contact manufacturer

Model or catalog number

BREEZE C3