This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

BOMImed Reusable Fiber Optic LED handle Med. and Small

Starting date:
January 11, 2013
Posting date:
February 4, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-17209

Products recalled

  1. Bomimed Reusable Fiber Optic LED handle Med. and Small

Reason

The inside diameter of the collar bushing that holds the LED bulb inside of the handle may potentially be too small to fit a replacement or different bulb. This may occur if a new or different bulb is used.

Affected products

A. BOMImed Reusable Fiber Optic LED handle Med. and Small

Lot or serial number
  • 2412
  • 2612
Model or catalog number
  • OL-324L7
  • OL-328L7
Companies
Manufacturer
Bomimed Inc.