This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

On-Board Imager (2014-03-20)

Starting date:
March 20, 2014
Posting date:
July 15, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40513

Recalled products

  1. On-Board Imager 

Reason

The shaft of the type 01 elbow motor for exact arms used in the on-board imager (OBI) device have failed due to belt over-tensioning.

Affected products

A. On-Board Imager 

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number
  • OBI
Companies
Manufacturer
Varian Medical Systems, Inc.
3100 Hansen Way
Palo Alto
94304
California
UNITED STATES