Health product recall

On-Board Imager (2014-03-20)

Starting date:
March 20, 2014
Posting date:
July 15, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40513

Recalled products

  1. On-Board Imager 

Reason

The shaft of the type 01 elbow motor for exact arms used in the on-board imager (OBI) device have failed due to belt over-tensioning.

Affected products

A. On-Board Imager 

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number
  • OBI
Companies
Manufacturer

Varian Medical Systems, Inc.

3100 Hansen Way

Palo Alto

94304

California