Blueprint Mixed Reality BP MxR Glenoid Pin Guide
Brand(s)
Last updated
Summary
Product
Blueprint Mixed Reality BP MxR Glenoid Pin Guide
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Blueprint Mixed Reality BP MxR Glenoid Pin Guide | 22A796 22A797 | MRUE200 |
Issue
During product complaint review (July 2024), Stryker has identified an issue that impacts specific lots of Blueprint® Mixed Reality Glenoid Pin Guides. For the impacted pins guides, the metallic tube may disassemble from the plastic handle. This can result in the Blueprint Mixed Reality headset from being able to detect the Blueprint Mixed Reality Glenoid Pin Guide.
Recall Start Date: November 26, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies
Tornier S.A.S.
161, Rue Lavoisier, Montbonnot-Saint-Martin, France, 38330
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76680
Get notified