Health product recall

Blueprint Mixed Reality BP MxR Glenoid Pin Guide

Brand(s)
Last updated

Summary

Product
Blueprint Mixed Reality BP MxR Glenoid Pin Guide
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Blueprint Mixed Reality BP MxR Glenoid Pin Guide 22A796 22A797 MRUE200

Issue

During product complaint review (July 2024), Stryker has identified an issue that impacts specific lots of Blueprint® Mixed Reality Glenoid Pin Guides. For the impacted pins guides, the metallic tube may disassemble from the plastic handle. This can result in the Blueprint Mixed Reality headset from being able to detect the Blueprint Mixed Reality Glenoid Pin Guide.

Recall Start Date: November 26, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies

Tornier S.A.S.

161, Rue Lavoisier, Montbonnot-Saint-Martin, France, 38330

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76680

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe