BLOOD MONITORING UNIT 40 (BMU 40) (2020-09-22)
- Starting date:
- September 22, 2020
- Posting date:
- October 16, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-74117
Last updated:
2020-10-15
Affected Products
BLOOD MONITORING UNIT 40 (BMU 40)
Reason
The printed circuit board (PCBA Connector) of the BMU40 was designed with a short creepage distance. This PCBA Connector is positioned in the inner and backside of the metal enclosure of the BMU40. This may lead to shock, loss of data, or delay of use. Maquet Cardiopulmonary GmbH has not received any reports of serious injuries or death due to the malfunction of the PCBA Connector.
Affected products
BLOOD MONITORING UNIT 40 (BMU 40)
Lot or serial number
90002058
90002102
90002103
90002107
90002111
Model or catalog number
BMU40
Companies
- Manufacturer
-
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY