Health product recall

BLANKETROL II & III System - Main Unit (2019-07-02)

Starting date:
July 2, 2019
Posting date:
August 9, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70647



Last updated: 2019-08-09

Affected Products

A. BLANKETROL III System - Main Unit

B. BLANKETROL II - Main Unit

Reason

The BLANKETROL manual does not provide instruction stating that exceeding 40°C for extended periods can cause tissue damage.

Affected products

A. BLANKETROL III System - Main Unit

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

233

Companies
Manufacturer

Cincinnati Sub-Zero Products

12011 Mosteller Road

Cincinnati

Ohio

UNITED STATES


B. BLANKETROL II - Main Unit

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

222R

222S

Companies
Manufacturer

Cincinnati Sub-Zero Products

12011 Mosteller Road

Cincinnati

Ohio

UNITED STATES