Health product recall

BIS Complete Monitoring System (2018-07-26)

Starting date:
July 26, 2018
Posting date:
August 17, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67386

Affected products

BIS Complete Monitoring System

Reason

Replacement instructions were revised for the Li-Ion battery packs used in BIS Vista Monitoring System. These revised instructions will require battery pack replacement after 4 years due to the potential for aged battery packs to experience an internal short that may result in a thermal event.

Affected products

BIS Complete Monitoring System

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

186-1046

Companies
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES