This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Bionic Comfortline Chair (2014-03-24)
- Starting date:
- March 24, 2014
- Posting date:
- April 14, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-39025
Recalled Products
Bionic Comfortline Chair
Reason
The instructions for use of the therapy chair - Comfortline Canada - serial number starting from 06C06-018 has been amended to included the following: During installation of the therapy chair the installing staff has to make sure that the seat of the chair is installed correctly and according to the instruction for use. If the seat is not locked properly it might slide unintentionally.
Affected products
Bionic Comfortline Chair
Lot or serial number
Serial numbers starting from 06C06-018
Model or catalog number
- 109434
- 110014
- 110015
- 113789
- 114665
- LOC-17039
- LOC-17070
- LOC-17097
- LOC-17119
- NVM-10336
Companies
- Manufacturer
-
Bionic Medizintechnik GmbH
Max-Planck-Strasse 21
Friedrichsdorf
61381
GERMANY