BIOHIT Pipette for easyMAG (2014-12-23)
- Starting date:
- December 23, 2014
- Posting date:
- January 14, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-43235
Recalled Products
BIOHIT Pipette for easyMAG
Reason
In some manufacturing lots, the top cover of the pipette AC-adaptor opens easily leaving the electrical parts visible and causing a potential risk of electrical shock and/or burning.
Affected products
BIOHit Pipette for easyMAG
Lot or serial number
2314 to 5013
Model or catalog number
280147
Companies
- Manufacturer
-
Biomerieux SA
chemin de l'Orme
Marcy l'Etoile
69280