Health product recall

BIOCAL 370 Temperature Controller (2018-02-08)

Starting date:
February 8, 2018
Posting date:
March 2, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66074

Affected products

BIOCAL 370 Temperature Controller

Reason

Safety issues were raised by regulators regarding water systems quality of temperature controllers, regardless of manufacturer. Medtronic specifically assessed the associated risk with the use of the Medtronic BIOCAL heater/cooler model 370. The potential hazard associated was identified as the presence of mycobacterium in the operating room due to the BIOCAL device not cleaned /disinfected on a regular basis, or the use of non- deionized water.

Affected products

BIOCAL 370 Temperature Controller

Lot or serial number

All lots.

Model or catalog number

95161
R95161-00

Companies
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES