BIOCAL 370 Temperature Controller (2018-02-08)
- Starting date:
- February 8, 2018
- Posting date:
- March 2, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66074
Affected products
BIOCAL 370 Temperature Controller
Reason
Safety issues were raised by regulators regarding water systems quality of temperature controllers, regardless of manufacturer. Medtronic specifically assessed the associated risk with the use of the Medtronic BIOCAL heater/cooler model 370. The potential hazard associated was identified as the presence of mycobacterium in the operating room due to the BIOCAL device not cleaned /disinfected on a regular basis, or the use of non- deionized water.
Affected products
BIOCAL 370 Temperature Controller
Lot or serial number
All lots.
Model or catalog number
95161
R95161-00
Companies
- Manufacturer
-
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES