BeneVision N1 Patient Monitor
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
BeneVision N1 Patient Monitor |
More than 10 numbers, contact manufacturer. |
BENEVISION N1 |
Issue
Mindray has identified that, under rare circumstances, using the BeneVision N1 Patient Monitor with software versions 01.60.00.01 and 01.61.00.01 as a multi-parameter module in conjunction with a host monitor may result in the activation of an abnormal alarm pause. This can occur if the alarm pause button is pressed twice rapidly on either the N1 or the host monitor.
Recall start date: August 19, 2025