Health product recall

Benchmark Ultra Automated Immunostaining System & Benchmark XT (2020-04-08)



Starting date:
April 8, 2020
Posting date:
May 8, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73029


Last updated:

Affected Products

  1. Benchmark Ultra Automated Immunostaining System
  2. Benchmark XT

Reason

Following the information provided by our manufacturer on investigation of customer complaints, we would like to inform you of an issue that has been identified in Ventana software system (VSS V 11.0 to 12.5.3 when used with identified staining instruments) that can result in the software allowing a dispenser to be used beyond their intended capacity when a slide is aborted/stopped while the SW is in compiling or starting state. The manufacturer advised us of this on 06-Apr-2020.

Affected products

A. Benchmark Ultra Automated Immunostaining System

Lot or serial number
  • More than 10 numbers, contact manufacturer.
Model or catalog number
  • N750-BMKU-FS
Companies
Manufacturer

Ventana Medical Systems Inc.

1910 E Innovation Park Drive,

Tucson

85755

Arizona

UNITED STATES



B. Benchmark XT

Lot or serial number
  • More than 10 numbers, contact manufacturer.
Model or catalog number
  • N750-BMKXT-FS
Companies
Manufacturer

Ventana Medical Systems Inc.

1910 E Innovation Park Drive,

Tucson

85755

Arizona

UNITED STATES