Benchmark ULTRA Automated Immunostaining System (2020-09-10)
- Starting date:
- September 10, 2020
- Posting date:
- September 25, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73999
Last updated:
2020-09-25
Affected Products
Benchmark ULTRA Automated Immunostaining System
Reason
Fluid leaks within BenchMark ULTRA instruments that created an electrical short inside the power socket, which over time generated heat, smoke, and eventually burning and melting of the electrical cord connector and socket.
Affected products
Benchmark ULTRA Automated Immunostaining System
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
256476
Companies
- Manufacturer
-
Ventana Medical Systems Inc.
1910 E Innovation Park Drive,
Tucson
85755
Arizona
UNITED STATES