Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73999


Last updated:

Affected Products

Benchmark ULTRA Automated Immunostaining System

Reason

Fluid leaks within BenchMark ULTRA instruments that created an electrical short inside the power socket, which over time generated heat, smoke, and eventually burning and melting of the electrical cord connector and socket.

Affected products

Benchmark ULTRA Automated Immunostaining System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

256476

Companies
Manufacturer

Ventana Medical Systems Inc.

1910 E Innovation Park Drive,

Tucson

85755

Arizona

UNITED STATES