This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
The Belmont Rapid Infuser, FMS2000
- Starting date:
- October 5, 2012
- Posting date:
- October 29, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices, Product Safety
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-15142
Recalled Products:
The Belmont Rapid Infuser, FMS2000
Reason
There is a potential that fluid spills may result in an accumulation of salt build up on the power entry area and may result in the remote potential of an electrical flash occurring.
Affected products
The Belmont Rapid Infuser, FMS2000
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 903-00001
- 903-00001A (230V)
- 903-00023 (120V)
Companies
- Manufacturer
- Belmont Instrument Corp.