Bellavista Ventilator
Brand(s)
Last updated
Summary
Product
Bellavista Ventilator
Issue
Medical devices - Premature failure
What to do
| Contact the manufacturer if you require additional information. |
Affected products
| Affected Products | Lot or serial number | Model or catalog number |
|---|---|---|
| Bellavista 1000 Ventilator | More than 10 numbers, contact manufacturer. | 301.100.000 |
| Bellavista 1000E Ventilator | More than 10 numbers, contact manufacturer. | 301.100.100 |
Issue
Ventilation stops during clinical use, with waveforms and parameters frozen (not updated). a continuous audible/visual high priority alarm for technical failure 305 is generated. ventilation is suspended until unit is rebooted/replaced.
Recall start date: 2021-12-20
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
| Imtmedical Ag |
| Gewerbestrasse 8, Buchs, Switzerland, 9470 |
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type I
Identification number
RA-63776