This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

BEHRING ELISA PROCESSOR SYSTEM- VARICELLA IGG ENZYGNOST (2015-10-08)

Starting date:
October 8, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55650

Affected Products

BEHRING ELISA PROCESSOR SYSTEM- VARICELLA IGG ENZYGNOST

Reason

Siemens Healthcare Diagnostics Products has confirmed that the mentioned Enzygnost Anti-VZV/IgG lots can show a higher variation in homogeneity. This can lead to invalid runs by not meeting the run validation criterion for the reference p/n at the start and end of a series.

Affected products

BEHRING ELISA PROCESSOR SYSTEM- VARICELLA IGG ENZYGNOST

Lot or serial number

44218

44309

44565

45077

Model or catalog number

OWLT15

Companies
Manufacturer

Siemens Healthcare Diagnostics Products Gmbh

Emil-Von-Behring-Str. 76

Marburg

35041

GERMANY