Beckman Coulter AU System - Liquid Urine Chemistry Calibrator (2020-02-05)
- Starting date:
- February 5, 2020
- Posting date:
- February 28, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72437
Last updated: 2020-03-06
Affected Products
Beckman Coulter AU System - Liquid Urine Chemistry Calibrator
Reason
Liquid Urine Chemistry Calibrator (lot 1809005B) may contain precipitate which when used with uric acid reagent ref OSR6X98 (OSR6098, OSR6198, OSR6298, AND OSR6698), may lead to incorrect calibration and QC failures.
Affected products
Beckman Coulter AU System - Liquid Urine Chemistry Calibrator
Lot or serial number
1809005B
Model or catalog number
DR0090