Health product recall

Beckman Coulter Au Immunochemistry System- Immunoglobulin M (IgM)

Last updated

Summary

Product
Beckman Coulter Au Immunochemistry System- Immunoglobulin M (IgM)
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Beckman Coulter Au 2700 Immunochemistry System - Immunoglobulin M (IgM)

All lots.

OSR61173

Beckman Coulter Au 5400 Immunochemistry System - Immunoglobulin M (IgM)

All lots.

OSR61173

Beckman Coulter Au Immunochemistry System- Immunoglobulin M (IgM)

All lots.

OSR61173

Issue

Beckman Coulter has determined that the LIH influence check for IgM lipemic interference on the AU480, AU680, DxC 500 AU and DxC 700 AU analyzers incorrectly states "+++++", which is equivalent to >500 mg/dL of intralipid. The correct LIH influence check for IgM lipemic interference for the AU480, AU680, DxC 500 AU and DxC 700 AU analyzers is "+++", which is equivalent to 200-299 mg/dL of intralipid. The LIH influence check for IgM lipemia interference for the AU5800 is not affected. The lipemia interference specification in the IgM Instructions For Use (IFU), BLOSR6X173 and BLOSR6X173EU is also not affected. 

Recall Start Date: May 27, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Beckman Coulter, Inc.

250 S. Kraemer Blvd., Brea, California, United States, 92821

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75639

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe