Beckman Coulter AU Immunochemistry System- HDL/LDL Cholesterol Assay
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue |
|---|---|---|
Beckman Coulter AU Immunochemistry System- HDL/LDL Cholesterol Assay |
More than 10 numbers, contact manufacturer. |
OSR6196 |
Beckman Coulter AU Immunochemistry System- HDL/LDL Cholesterol Assay |
More than 10 numbers, contact manufacturer. |
OSR6296 |
Issue
LDL cholesterol OSR6x96 does not meet the triglyceride interference claim per the BAOSR6x96 v10 IFU which states "no significant interference up to 1,500 mg/dl triglyceride" where the criteria for no significant interference is recovery within 10% of the initial value as outlined in the IFU.