Beckman Coulter AU Immunochemistry System - Bicarbonate (CO2) Assay
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Beckman Coulter AU Immunochemistry System - Bicarbonate (CO2) Assay |
More than 10 numbers, contact manufacturer. |
OSR6237 |
Beckman Coulter AU Immunochemistry System - Bicarbonate (CO2) Assay |
More than 10 numbers, contact manufacturer. |
OSR6137 |
Issue
Customer complaints were received for multiple bicarbonate reagent lots associated with false high patient results for bicarbonate. The high results are due to elevated levels of LDH in the patient samples tested. The current bicarbonate product IFU's do not advise of LDH interference with this assay.
Additional information
Details
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821
Get notified
Receive emails about new and updated recall and safety alerts.