Health product recall

Beckman Coulter Au 2700/5400 Immunochemistry System - Immunoglobulin A (IgA)

Last updated

Summary

Product
Beckman Coulter Au 2700/5400 Immunochemistry System - Immunoglobulin A (IgA)
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Beckman Coulter Au 5400 Immunochemistry System - Immunoglobulin A (IgA) All lots. OSR61171
Beckman Coulter Au2700 Immunochemistry System - Immunoglobulin A (IgA) All lots. OSR61171

Issue

Internal testing determined that IgA OSR61171 failed to meet the interference claim as stated per the IFU for the lipemic serum interference. All lots of IgA OSR61171 are affected.

Recall start date: July 17, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Beckman Coulter, Inc.

250 S. Kraemer Blvd., Brea, California, United States, 92821

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74059

Get notified

Receive notifications for new and updated recalls and alerts by category.