Product
Bd Veritor Plus Analyzer
Issue
Medical devices - Increased risk or new adverse events
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Bd Veritor Plus Analyzer |
More than 10 numbers, contact manufacturer. |
256066 |
Issue
Bd has received two (2) complaints reporting that the Bd Veritor Plus analyzer may have the potential to overheat and/or cause fire. Bd's investigation has found that the issue can occur when an unauthorized alternative ac power adapter, and not the one provided by Bd with the Bd Veritor Plus analyzer, is used to power the instrument.
Recall start date: Jan 5, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Companies
| Becton Dickinson And Company (Bd) |
|
7 Loveton Circle, Sparks, Maryland United States, 21152 |
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63828