Health product recall

BD Vacutainer Heparin and EDTA Tubes (2018-04-06)

Starting date:
April 6, 2018
Posting date:
May 4, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66698

Affected products

A. BD Vacutainer Heparin Tubes

B. BD Vacutainer EDTA Tubes

Reason

There is the potential for low bias (underestimation of blood lead levels) that can lead to delayed or inappropriate, false negative, test results, when using the BD Vacutainer EDTA (Lavender, Tan, and Pink Top) tubes and BD Vacutainer Heparin (Green Top) tubes with the Magellan Diagnostics Leadcare Testing Systems.

Affected products

A. BD Vacutainer Heparin Tubes

Lot or serial number

All Lots within Expiry

Model or catalog number

366664

367884

367886

Companies
Manufacturer

Becton Dickinson and Company

1 Becton Drive

Franklin Lakes

07417

UNITED STATES



B. BD Vacutainer EDTA Tubes

Lot or serial number

All Lots within Expiry

Model or catalog number

366401

366643

367841

367844

367855

367856

367861

367863

367899

Companies
Manufacturer

Becton Dickinson and Company

1 Becton Drive

Franklin Lakes

07417

UNITED STATES