Health product recall

BD Vacutainer Eclipse Blood Collection Needle (2019-12-11)



Starting date:
December 11, 2019
Posting date:
January 6, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72009


Last updated:

Affected Products

BD Vacutainer Eclipse Blood Collection Needle

Reason

Absence of bevel on the non-patient (NP) needle end of the Eclipse Blood Collection Needle product causing blood leakage. If there is no bevel, the needle may damage the sleeve, causing leakage. Additionally, it may make it difficult to insert BD Vacutainer products onto the NP end.

Affected products

BD Vacutainer Eclipse Blood Collection Needle

Lot or serial number

9010765

Model or catalog number

368608

Companies
Manufacturer

Becton Dickinson and Company

1 Becton Drive

Franklin Lakes

07417

New Jersey

UNITED STATES