BD Synapsys™ Informatics Solution
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
BD Synapsys™ Informatics Solution |
Version # 4.20,4.30,5.10,5.20,5.30 |
444150 |
Issue
BD identified through one (1) complaint that for BD Kiestra™ customers with BD Synapsys™, running Synapsys™ software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys™ will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint.
Recall start date: September 12, 2024
Additional information
Details
Becton Dickinson And Company (BD)
7 Loveton Circle, Sparks, Maryland, United States, 21152
Get notified