Health product recall

BD Synapsys

Last updated

Summary

Product
BD Synapsys
Issue
Medical devices - Performance
What to do

Review recall notice, share/post within facility, complete/return reply form, scheddule remote servicing. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
BD Synapsys Version 2.1 and higher 444150

Issue

BD recently identified, through a customer complaint investigation, that there is a risk of mis-association of patient demographic data for BD Synapsys™ users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (skip, final discard, mark as read, save) to process patient results rapidly.

Recall start date : June 8, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Becton Dickinson And Company (BD)

7 Loveton Circle, Sparks, Maryland, United States, 21152

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73872

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe