BD SafetyGlide 3ml 25G x 1 (2020-07-29)
- Starting date:
- July 29, 2020
- Posting date:
- August 14, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73713
Last updated:
Affected Products
BD SafetyGlide 3ml 25G x 1
Reason
Printed scale markings on the 3ml SafetyGlide syringes, with catalogue # 305924 from batch # 8289599, are printed incorrectly. The scale is skewed to different degrees, resulting in missing and/or partially printed scale numbers and scale lines.
Affected products
BD SafetyGlide 3ml 25G x 1
Lot or serial number
8289599
Model or catalog number
305924
Companies
- Manufacturer
-
Becton Dickinson and Company
1 Becton Drive
Franklin Lakes
07417
UNITED STATES