BD Phoenix System - NMIC 1 and NMIC 93 (2018-01-09)
- Starting date:
- January 9, 2018
- Posting date:
- February 1, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65880
Affected products
- BD PHOENIX SYSTEM - NMIC 1
- BD PHOENIX NMIC-93
Reason
There is an increased rate of very major errors (vme) for Colistin with Colistin resistant strains of the Enterobacteriaceae (17.5%). This percentage is derived from a recently-concluded external study, and likely involves emerging hetero-resistant strains..
Affected products
A. BD PHOENIX SYSTEM - NMIC 1
Lot or serial number
- 7095829
- 7276978
Model or catalog number
448063
Companies
- Manufacturer
-
Becton Dickinson and Company
7 Loveton Circle
Sparks
21152
Maryland
UNITED STATES
B. BD PHOENIX NMIC-93
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 448764
- 448788
Companies
- Manufacturer
-
Becton Dickinson and Company
7 Loveton Circle
Sparks
21152
Maryland
UNITED STATES