BD FACSLyric Flow Cytometer
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
BD FACSLyric Flow Cytometer |
More than 10 numbers, contact manufacturer. |
659180 |
Issue
BD has identified through complaints an increase in field failures / replacements of the power supply module on BD FACSLyric Flow Cytometers. The capacitor on the power supply can split which has the potential for the instruments to fail to power on and/or stay on.
Recall Start Date: November 20, 2024
Additional information
Details
Becton, Dickinson And Company Bd Biosciences
155 North Mccarthy Boulevard, Milpitas, California, United States, 95035
Get notified
Receive emails about new and updated recall and safety alerts.