Health product recall

BD BBL Gram Stain Kit & Crystal Violet (2020-04-20)



Starting date:
April 20, 2020
Posting date:
May 8, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73025


Last updated:

Affected Products

  1. BD BBL Gram Stain Kit
  2. BD BBL Gram Crystal Violet

Reason

Confirmed customer reports that the products may exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci. Use of products with increased levels of artifacts could result in a false-positive gram stain. Gram stain results must be confirmed with additional procedures, but if these protocols are not followed, a false-positive could lead to administration of unnecessary antibiotics or surgical procedures.

Affected products

A. BD BBL Gram Stain Kit

Lot or serial number
  • 7255
  • 7261
  • 14322
  • 411114
Model or catalog number
  • 212539
Companies
Manufacturer

Becton Dickinson and Company

7 Loveton Circle

Sparks

21152

Maryland

UNITED STATES



B. BD BBL Gram Crystal Violet

Lot or serial number
  • 27824
  • 9269800
  • 9296589
  • 9296631
  • 9317733
  • 9352484
  • 9352492
Model or catalog number
  • 212525
  • 212526
Companies
Manufacturer

Becton Dickinson and Company

7 Loveton Circle

Sparks

21152

Maryland

UNITED STATES