Health product recall

BD BACTEC MGIT 960 PZA Kit And Susceptibility Test Medium

Last updated

Summary

Product
BD BACTEC MGIT 960 PZA Kit And Susceptibility Test Medium
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

BD BACTEC MGIT 960 PZA Kit And Susceptibility Test Medium

More than 10 numbers, contact manufacturer.

245128

Issue

Through internal BD complaint trending and subsequent internal raw material testing it was identified that the BD BACTEC™ MGIT™ 960 PZA Kits, listed in attachment 1 may intermittently produce falsely resistant results for pyrazinamide (PZA) during susceptibility testing of mycobacterium tuberculosis isolates. 

Recall start date: July 18, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Microbiology
Companies
Becton Dickinson And Company (Bd)
7 Loveton Circle, Sparks, Maryland, United States, 21152
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75886

Get notified

Receive notifications for new and updated recalls and alerts by category.