BD BACTEC MGIT 960 PZA Kit And Susceptibility Test Medium
Brand(s)
Last updated
Summary
Product
BD BACTEC MGIT 960 PZA Kit And Susceptibility Test Medium
Issue
Medical devices - Sterility
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
BD BACTEC MGIT 960 PZA Kit And Susceptibility Test Medium |
More than 10 numbers, contact manufacturer. |
245128 |
Issue
Through internal BD complaint trending and subsequent internal raw material testing it was identified that the BD BACTEC™ MGIT™ 960 PZA Kits, listed in attachment 1 may intermittently produce falsely resistant results for pyrazinamide (PZA) during susceptibility testing of mycobacterium tuberculosis isolates.
Recall start date: July 18, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Microbiology
Companies
Becton Dickinson And Company (Bd) |
7 Loveton Circle, Sparks, Maryland, United States, 21152 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75886
Get notified