Health product recall

BBL™ Chocolate II Agar

Last updated

Summary

Product
BBL™ Chocolate II Agar
Issue
Medical devices - Performance
What to do

Destroy remaining inventory, share recall notice with users, complete/return response form. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

BBL™ Chocolate II Agar

2206461

221267

Issue

BD has confirmed through complaint investigation that a single lot of BBL™ Chocolate II Agar has misformulated nicotinamide adenine dinucleotide (NAD) concentration in a portion of the batch. Specifically affected are box numbers 0002 through 0097. These boxes correlate to timestamps on the plates of 11:34 through 13:51. All other box numbers and plate stamps are unaffected and perform as intended.

Recall start date: November 21, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Becton Dickinson And Company (BD)

7 Loveton Circle, Sparks, Maryland, United States, 21152

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72233

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe