BBL MGIT Mycobacteria Growth Indicator Tube (2013-09-09)
- Starting date:
- September 9, 2013
- Posting date:
- October 17, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-36231
Recalled products
BBL MGIT Mycobacteria Growth Indicator Tube
Reason
Products have been recalled for the following labelling issues.
Product contained incorrect barcodes where the labels began with ""43 02"" rather than the required ""43 01"". These batches were placed on quality notification and dispositioned for rework to over-label the barcode area with the correct symbology. The over-label was manually applied to each tube.
One customer complaint was received where the over-label had partially peeled off of the tube, the incorrect barcode underneath was scanned into the MGIT instrument, and the epicenter connected to the customer's laboratory system incurred an error. The customer also noted that they found additional tubes in their shipment that had multiple labels or no over-label.
The label prefix, specifically the 3rd and 4th digits that indicate the media type, gives the instrument information for what algorithm to apply for growth detection. Growth detection testing will be performed by the MGIT instrument every 60 minutes as usual on tubes containing incorrect barcode prefixes, but will not flag as positive when growth is visible because media type "43 02" is not defined. A barcode overlabel peeling off, exposing an incorrect barcode, causes MGIT false negative results by not flagging positive tubes. Susceptibility tests are not affected since susceptibility testing is conducted in a carrier where instrument instruction is provided by the carrier.
Affected products
BBL MGIT Mycobacteria Growth Indicator Tube
Lot or serial number
- 3053240
- 3061013
Model or catalog number
- 245122