Health product recall

Baxter Dialysis Machines (2020-11-17)



Starting date:
November 17, 2020
Posting date:
December 11, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74511


Last updated:

Affected Products

A. AK 95 S

B. AK 200 S

C. AK 200 ULTRA S

D. AK 96

E. AK 96

F. AK 98

G. AK 98

Reason

Baxter has not validated the use of any connectors placed between the blood set and the patient's blood access with Baxter dialysis machines. The use of connectors with potentially incompatible material may increase the risk for leakage in the extracorporeal circuit and may prevent a secure connection between the blood set and the patient's blood access device.

Affected products

A. AK 95 S

Lot or serial number

All lots.

Model or catalog number

AK-95S

Companies
Manufacturer

Gambro Lundia AB

MAGISTRATSVAGEN 16, P.O. BOX 10101,

Lund

22010

SWEDEN



B. AK 200 S

Lot or serial number

All lots.

Model or catalog number

107470

Companies
Manufacturer

Gambro Lundia AB

MAGISTRATSVAGEN 16, P.O. BOX 10101,

Lund

22010

SWEDEN



C. AK 200 ULTRA S

Lot or serial number

All lots.

Model or catalog number

107469

Companies
Manufacturer

Gambro Lundia AB

MAGISTRATSVAGEN 16, P.O. BOX 10101,

Lund

22010

SWEDEN



D. AK 96

Lot or serial number

All lots.

Model or catalog number

110661

Companies
Manufacturer

Gambro Lundia AB

MAGISTRATSVAGEN 16, P.O. BOX 10101,

Lund

22010

SWEDEN



E. AK 96

Lot or serial number

All lots.

Model or catalog number

110373

Companies
Manufacturer

Gambro Lundia AB

MAGISTRATSVAGEN 16, P.O. BOX 10101,

Lund

22010

SWEDEN



F. AK 98

Lot or serial number

All lots.

Model or catalog number

115249

115251

Companies
Manufacturer

Gambro Lundia AB

MAGISTRATSVAGEN 16, P.O. BOX 10101,

Lund

22010

SWEDEN



G. AK 98

Lot or serial number

All lots.

Model or catalog number

955406

955407

Companies
Manufacturer

Gambro Lundia AB

MAGISTRATSVAGEN 16, P.O. BOX 10101,

Lund

22010

SWEDEN