Bard Mission Disposable Core Biopsy Instrument Kit
Brand(s)
Summary
Review notice/ share with users, identify/destroy remaining inventory, complete/return response form.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Bard Mission Disposable Core Biopsy Instrument Kit |
More than 10 numbers, contact manufacturer. |
1820MSK |
Issue
BD continues to receive complaints for the product regarding mismatch between the coaxial and the needle in mission kit devices. Based on the event reported, the internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue.
Recall start date: November 8, 2023
Additional information
Details
| Bard Peripheral Vascular, Inc. |
| 1625 West 3rd Street, Tempe, Arizona, United States, 85281 |
Get notified