Health product recall

Bard Mission Disposable Core Biopsy Instrument Kit

Last updated

Summary

Product
Bard Mission Disposable Core Biopsy Instrument Kit
Issue
Medical devices - Device compatibility
What to do

Review notice/share with users, identify/destroy remaining inventory, complete/return response form. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Bard Mission Disposable Core Biopsy Instrument Kit

0001445630
0001451625
0001449980
0001450413
0001453517

1816MSK
1610MSK

Issue

BD has received complaints regarding mismatch between the coaxial and the needle in mission kit devices, where the external diameter of the biopsy instrument being larger than the internal diameter of the coaxial needle.

Recall start date: February 28, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies

Bard Peripheral Vascular, Inc.

1625 West 3rd Street, Tempe, Arizona, United States, 85281

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73130

Get notified

Receive emails about new and updated recall and safety alerts.