Health product recall

Bard Marquee Disposable Core Biopsy Instrument Kit

Last updated

Summary

Product
Bard Marquee Disposable Core Biopsy Instrument Kit
Issue
Medical devices - Performance
What to do

Review notice & share with users, discontinue/dispose affected inventory, complete/return reply form. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Bard Marquee Disposable Core Biopsy Instrument Kit

0001480774

MQK2016

Issue

BD has received complaints regarding mismatch between the coaxial and the needle in Bard® Marquee® kit devices, where the coaxial cannula inner diameter prohibits the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue.

Recall Start Date: August 10, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies

Bard Peripheral Vascular, Inc.

1625 West 3rd Street, Tempe, Arizona, United States, 85281

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74142

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe