BacT/Alert 3D and 3D Combo Systems
- Starting date:
- March 1, 2013
- Posting date:
- June 17, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-34193
Recalled products
- BacT/Alert 3D Combo
- BacT/Alert 3D
Reason
Due to a manufacturing error, the potential exists for a BacTAlert drawer to not close completely or to open during instrument operation, causing a possible decrease in incubator temperature.
Affected products
A. BacT/Alert 3D Combo
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
- 200290
- 200291