This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

BacT/Alert 3D and 3D Combo Systems

Starting date:
March 1, 2013
Posting date:
June 17, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-34193

Recalled products

  1. BacT/Alert 3D Combo
  2. BacT/Alert 3D

Reason

Due to a manufacturing error, the potential exists for a BacTAlert drawer to not close completely or to open during instrument operation, causing a possible decrease in incubator temperature.

Affected products

A. BacT/Alert 3D Combo

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number
  • 200290
  • 200291
Companies
Manufacturer

Biomerieux Inc.

595 Anglum Road

Hazelwood

84894

Missouri

UNITED STATES



B. BacT/Alert 3D

Lot or serial number

More than 10 numbers, contact manufacturer 

Model or catalog number
  • 210159
  • 210161
Companies
Manufacturer

Biomerieux Inc.

595 Anglum Road

Hazelwood

84894

Missouri

UNITED STATES